{"id":"arxiv/2601.16261","url":"https://arxiv.org/abs/2601.16261","tags":["physics.ins-det","quant-ph"],"score":0,"title":"A First Demonstration of the SQUAT Detector Architecture: Direct Measurement of Resonator-Free Charge-Sensitive Transmons","domain":"physics.ins-det","source":"arxiv","authors":["H. Magoon","T. Aralis","T. Dyson","J. Anczarski","D. Baxter","G. Bratrud","R. Carpenter","S. Condon","A. Droster","E. Figueroa-Feliciano","C. W. Fink","S. Harvey","A. Simchony","Z. J. Smith","S. Stevens","N. Tabassum","B. A. Young","C. P. Salemi","K. Stifter","D. I. Schuster","N. A. Kurinsky"],"reviews":0,"subject":"physics.ins-det","summary":"The Superconducting Quasiparticle-Amplifying Transmon (SQUAT) is a new sensor architecture for THz (meV) detection based on a weakly charge-sensitive transmon directly coupled to a transmission line. In such devices, energy depositions break Cooper pairs in the qubit capacitor islands, generating quasiparticles. Quasiparticles that tunnel across the Josephson junction change the transmon qubit parity, generating a measurable signal. In this paper, we present the design of first-generation SQUATs and demonstrate an architecture validation. We summarize initial characterization measurements…","category":"research","comments":0,"disputed":false,"division":"physics","mentions":0,"discussed":0,"published":"2026-01-22","reactions":{}}